- Home
- Products
- Open frame Single & Multi-output
- MPM-S050 Series
MPM-S050 Series








FEATURES
○ Meets EMI CISPR 11 / FCC Part 18 class B
○ 60W air cooling/50W convection-cooled @ 50°C ambient
○ High efficiency up to 88%
○ No-load power consumption < 0.3W
○ Wide operating temperature -20~80°C
○ Compact size 2x3"
○ Medical standard IEC, EN, UL 60601- 1, type BF rated patient contact leakage current
Models & Ratings
Model Number | Wattage | Output Voltage | Output Current |
---|---|---|---|
MPM-S052 | 30/35W | +5V | 7A |
MPM-S053 | 50/60W | +12V | 5A |
MPM-S055 | 50/60W | +24V | 2.5A |
MPM-S056 | 50/60W | +48V | 1.25A |
Summary
Characteristic | Minimum | Typical | Maximum | Units | Notes & Conditions |
---|---|---|---|---|---|
Input Range | 85 | 115 / 230 | 264 | VAC | Universal input range |
Input Frequency | 47 | 50 / 60 | 63 | Hz | AC input |
Efficiency | 88 | % | At input 230VAC, rated load, 1.0 hr. warm up | ||
Operation Temperature | -20 | +80 | °C | Derate from 50°C, become 50% load at 80°C | |
Weight | 103.5 | g | |||
Dimensions | 76.2 (L) x 50.8 (W) x 25.6 (H) mm, Tolerance +/- 0.5mm | ||||
EMC | EN 55011, EN 60601-1-2, EN 61000-3-2, EN 61000-3-3, IEC 61000-4-2, IEC 61000-4-3, IEC 61000-4-4, IEC 61000-4-5, IEC 61000-4-6, IEC 61000-4-8, IEC 61000-4-11 | ||||
Safety Approvals | IEC 60601-1, EN 60601-1, ANSI/AAMI ES60601-1, CAN/CSA-C22. 2 No. 60601-1 |